Bioer offers three different types of products: no internal control, exogenous internal control, and endogenous internal control.
1. Reproducibility
QuantGene 9600 (96C)
FAM
HEX
CT value of the positive control diluted by 100 times
32.92
34.29
32.92
34.82
32.91
34.64
33.08
34.12
33.00
34.30
32.95
34.84
33.18
34.94
32.93
34.70
33.00
34.66
32.99
34.11
Mean Value
32.99
34.54
Standard Deviation
0.085
0.309
CV%
0.259%
0.895%
2. Low Concentration Sample Detection Rate
Samples
1000-fold dilution of positive serum
2000-fold dilution of positive serum
4000-fold dilution of positive serum
Channel
FAM
HEX
FAM
HEX
FAM
HEX
CT Value
38.86
28.34
38.82
28.32
40.39
28.49
38.70
28.06
38.59
28.05
39.91
28.22
38.83
28.22
38.06
28.05
40.10
28.18
39.35
28.25
38.97
27.92
41.17
28.40
37.48
27.98
38.28
27.93
40.07
28.17
38.32
28.07
37.64
28.00
/
29.53
38.15
28.02
38.69
28.08
/
28.42
37.83
28.03
38.66
28.17
39.32
28.21
Summary: In this validation, the detection rate for positive serum diluted by 2000 times is 100%, and the detection rate for a 4000-fold dilution is 75%.
1. The kit need to be transported under freezing conditions.
2. The kit should be stored at -25℃~-15℃ away from light, and avoid repeated freeze-thaw more than 5 times.
3. The kit can be stored for up to 12 months if all components are kept in the manner above. (Please use the kit in the period of validity).
Cat# | Description | Packing size |
BSL04M1A | African Swine Fever Virus Real Time PCR Detection Kit, with Internal Control | 50T |
BSL04M1B | African Swine Fever Virus Real Time PCR Detection Kit, No Internal Control | 50T |
BSL04M1 | African Swine Fever Virus Real Time PCR Detection Kit, With Exogenous Control | 50T |