Case 1. Accuracy: The positive reference products P1-P11 and the negative reference products N1-N10 of the reference products of Hangzhou Bioer Technology Co., Ltd. were extracted using the Bioer extraction reagent BSC86S1E (Zhehang Machinery Equipment No. 20200872) according to the instructions.
Result: The results showed that the adenovirus and adenovirus types 3, 4 and 7 could be detected accurately. The test results of enterprise reference products: the positive coincidence rate is 100%, and the negative coincidence rate is 100%.
Caese 2. Precision: After dissolving the precision reference products J1-J4 of the adenovirus nucleic acid and 3/4/7 typing detection kit (fluorescent PCR method) according to the requirements of use, the reagent BSC86S1E (Zhehang Machinery Equipment 20200872) was extracted by Biori No.) after extraction, three batches of reagents were used for detection, and the detection was repeated 10 times.
Result: The results showed that the precision variation coefficients of the three batches of reagents were all less than 5%, and the reagent precision was good.